An ion mobility spectrometer(IMS) equipped with a laser ionization source is used for the sensitive detection of formaldehyde wnder atmospheric pressure.Experimental apparatus is introduced.We also discussed the ion signal intensity of analyte wnder various drift voltages,different temperature and laser energy.Experimental results show that this method is very effective.and it has high sensitivity.Detection limits are determined to be in the upper ppb range.