Under the elaborate survey of recent development of the seawater salinity measurement technology,we systematically proposed a measurement principle,based on differential refractive-index and PSD (Position Sensitive Device),and built a mathematical model,designed the structure of sensing probe,produced a prototype.On the basis of numerous experimental data,we determined the salinity calibration curve,the system stability and interfering factors were also measured.Experimental results indicate the salinity re...