    摘要:

    形变图像配准(deformable image registration, DIR)是自适应放射治疗(adaptive radiotherapy, ART)中轮廓推衍和剂量累积的关键过 程。 本研究比较3种不同形变配准算法在自适应放疗过程中轮廓推衍和剂量累积的变化。选择头 颈部癌症患者的放疗影像数据图像进行这项研究,刚性配准后采用B样条弹性配准和LCC- Demons配准算法,其中B样条弹性算法研究了两种优化指标:归一化互相关、归一化互相 关结合弯曲能量惩罚(bending energy penalty, BEP) 正则化。剂量评估采用剂量参数D95、Dmin、Dmean、Dmax和均 匀性指数(homogeneity index, HI)来评价累积剂量与初始剂量的差异。几何评估使用Dice 相似性系数(Dice similarity coefficient, DSC)和 Hausdorff距离(Hausdorff distance, HD)来评价体积之间的重叠度。经过实验,几何评估所得的PTV(planning target volume)靶区、左腮腺、 右腮腺Dice系数在B样条弹性配准的两种优化算法和LCC-Demons算法中分别为:0. 82、 0.83、0.86,其中在LCC-Demons算法下得到的Dice系数达到了最大值0.91, 所得的Hausdorff距离均在1—3 mm之间,DSC重叠度结果与表明偏差 距离均在临床可接受的 范围内。对于剂量评估,相应的HI值分别为:1.089、1.082、1.081,LCC-Demons同样达到 了最小值1.081,本实验中LCC-Demons在轮廓推衍和剂量映射中均取 得了较好的效果,同 时也为临床环境下对配准质量的评估提供了一定的借鉴。

    Abstract:

    Deformation image registration (DIR) is a key process of contour derivation and dose accumulation in adaptive radiotherapy (ART).This study compares the changes of contour derivation and dose accumulation in the process of adaptive radiotherapy with th ree different deformation registration algorithms.The radiotherapy imaging data images of pat ients with head and neck cancer were selected for this study.After rigid registration,B-splin e elastic registration and LCC-Demons registration algorithms were used.The B-spline elastic algorit hm studied two optimization indexes:normalization Cross-correlation,normalized cross-correl ation combined with bending energy penalty (BEP) regularization.Dose evaluation uses dose parameters D95,Dmin, Dmean,Dmax and unifo rmity index (HI) to evaluate the difference between cumulative dose and initial dose.Geometric evaluation uses Dice correlation coefficient (DCS) and H ausdorff distance (HD) to evaluate the degree of overlap between volumes.After experiments,the Dice coefficients of the planning target volume (PTV) target area,left parotid gland and right parotid gland obtained by geometri c evaluation are respectively 0.82±0.07,0.83±0.07,0.86±0.08in the two optimization algorith ms of B-spline elastic registration and the LCC-Demons algorithm.Among them,the Dice coeffic ient obtained under the LCC-Demons algorithm reached the maximum value of 0.91, and the obtai ned Hausdorff distances were all between 1and 3mm.The DSC overlap result and the indicated deviation distance were both within the clinically acceptable range.For dose evaluation,the corr esponding HI values are:1.089,1.082,1.081,LCC-Demons also reached the minimum value of 1.081.I n this experiment, LCC-demons achieved good results in contour derivation and dose mapping,and it is also clinical it provides a certain reference for the evaluation of registration quality in th e environment.

    参考文献
    相似文献
    引证文献
引用本文

张雪洋,王玉,王明泉,张鹏程,檀桢.不同形变配准对头颈部CT图像的剂量评估研究[J].光电子激光,2022,33(2):157~162

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2020-01-09
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2022-03-24
  • 出版日期:
文章二维码